- Research
- Open access
- Published:
A soft wearable exoglove for rehabilitation assistance: a novel application of knitted shape-memory alloy as a flexible actuator
Fashion and Textiles volume 11, Article number: 14 (2024)
Abstract
We developed the knitted shape-memory alloy (SMA) as a flexible actuator and applied it to a soft wearable exoglove for rehabilitation assistance. Based on user needs and anthropometric data, we custom-designed the exoglove for three hemiplegic patients using the knitted-SMA actuator and evaluated its bending performance, gripping force and wearability under the four simulation conditions (S1–S4). To address the specific needs, both SMA plain- and double-knit modules were applied to the exoglove based on the patients’ finger joint range of motion (ROMs). The joint ROMs of all fingers increased by 13.71% and the skin temperature increased by 2.21 °C after actuating the glove (p = 0.006). The gripping force increased as much as 55.01%, when compared to the baseline. All patients were able to don and doff the developed glove independently, and positively evaluated their subjective satisfaction and thermal perception. The findings suggested the potential of the knitted SMA for the future development of soft wearable robots.
Introduction
Along with the global increase in the number of older people, the aging of the disabled population is emerging as an important issue in modern society. As of 2024, more than 46% of the global population aged 60 years or older have at least one disability, and this trend is predicted to increase (United Nations, 2022). Furthermore, approximately 16% of the global population currently suffers from severe disabilities that affect their daily lives (World Health Organization, 2023). In particular, hemiplegia caused by brain lesions results in abnormal muscle activity, skin reflexes, and spasticity owing to muscle paralysis, which makes it challenging to control voluntary movements (Cauraugh et al., 2000). The hands of hemiplegic patients, which have the most complex anatomical structure in the human body, can experience difficulties during daily activities such as grabbing objects, writing, and button fastening because their function is weakened (Nasir & Troynikov, 2019). Functional impairment in the hands of hemiplegic patients is caused by decreased activity of the extensor muscles of the fingers and the inappropriate simultaneous activation of the flexor muscles, and more motor defects appear in the extension function than the flexion function (Cauraugh et al., 2000).
Most treatment strategies for hemiplegic disorders begin by reducing spasticity (Ji et al., 2013). Thermal stimulation in localized parts of the body can diminish pain in paralyzed areas by increasing the blood circulation in the body’s internal tissues to relieve muscle stiffness (Kim et al., 2005). In addition, repetitive hand flexion and extension exercises can improve hand function by improving limb sensation and reducing muscle tension in hemiplegic patients (Kawahira et al., 2010). Thus, continuous and steady exercise with thermal stimulation is effective in increasing the joint range of motion (ROM) (Matsumoto et al., 2010). Furthermore, people with disabilities who use rehabilitation devices tend to have higher functional-recovery efficiency than those who only receive physical therapy (Yoon et al., 2018).
Recently, research has been conducted on assistive devices that are necessary to improve the life quality of people with disabilities. However, actual user satisfaction and usability of assistive devices are low, owing to the high cost of rehabilitation treatment and low wearability of the devices (Yeung et al., 2016). In particular, robotic assistive devices such as exoskeletons cause discomfort because of their large size and low flexibility, and they tend to focus only on the function of assisting daily activities rather than relieving the stiffness that is typically experienced by hemiplegic patients (Cappello et al., 2018; Kang et al., 2019). Furthermore, if people with physical disabilities wear ill-fitting assistive devices, they may stop using them, thereby furthering functional loss and mental-health problems (Yeung et al., 2016). Wearable assistive devices go beyond the concept of attaching assistive devices to the human body; in other words, they are worn on the body and can interact with each other. To increase user satisfaction, it is necessary to design assistive devices that are lightweight and flexible, have good wearability, and minimize the burden on the human body when applied to a body part (Walsh, 2018).
Toward this end, researchers have adopted shape-memory alloys (SMAs) as a flexible smart material, for its unique characteristic of returning to a memorized shape (Granberry et al., 2019). When the SMA wire is structured in a knit form, it can be applied to the body curvature at synovial joints as it can provide not only one-dimensional contraction deformation but also bending and contraction changes; further, it can provide a higher driving force when using a shorter wire (Han & Ahn, 2017). Lightweight and flexible actuators can be made without additional materials through the inherent flexibility of weft knit with continuity of loops and shape-transformation characteristics of SMA, making the smart material suitable for human-oriented assistive devices that can be applied to various body motions.
Therefore, we aimed to develop a rehabilitation-assistive exoglove for hemiplegic patients with the use of a knitted SMA. To achieve this goal, we took the ergonomic wearable design approach, which was characterized by a two-way interaction between the human body and wearable device, with an ultimate intention to optimize the desired functionality and positive user experience. As the first step, we identified specific user needs using the FEA Consumer Needs Model (Lamb & Kallal, 1992), and brainstormed a glove design solution based on the identified user needs. Next, we analyzed the changes in joint ROMs in the recruited hemiplegic patients to gauge the required mobility range for the rehabilitation-assistive exoglove, which also informed the bending and contraction span of the knitted SMA module. Once the exoglove prototype was developed, we evaluated the glove’s and assessed its wearability with the actual human subjects.
Literature Review
Soft wearable robotics
Soft wearable robotics is an emerging field that can solve the limitations of heavy and rigid materials, which can burden the human body when using exoskeleton robots. These robots were developed to enhance the performance of physical activities, such as helping healthy people lift heavy loads, reducing energy consumption during running, and assisting in movements for patients with reduced muscle strength (Walsh, 2018). Quinlivan et. al. (2017) developed a fabric-based multi-jointed soft exosuit that reduced the metabolic rate by approximately 20% by assisting the ankle joint during walking. Kim et. al. (2022) developed an exosuit that was capable of assisting hip flexion and reducing energy consumption during walking by approximately 15%. In et. al. (2015) developed a wearable robot for the hand that could help grasp various objects using a soft tendon routing system. In addition, tendon-driven soft wearable gloves were developed to assist finger movements using SMA wire (Hadi et al., 2018; Wang et al., 2021). Recently, soft wearables have been developed in a form similar to clothing, and interest in their functionality and wearability is increasing (Walsh, 2018). A soft exoskeleton provides the advantages of being supported by the wearer’s skeletal structure and conforming to the user’s body better than a rigid exoskeleton can.
Rehabilitation-assistive technology
A key component of assistive technology for people with disabilities is rehabilitation, which is a comprehensive concept that includes rehabilitation products and related services covering all types of disabilities (De Witte et al., 2018). These rehabilitation-assistive technologies, particularly those for hands, primarily enable motion in stiff fingers to improve the quality of daily living. Recent attempts have been made to develop assistive devices using lightweight and flexible textile materials with good wearability. Zhang et. al. (2017) developed a ring-type assistive device using a pneumatic-based actuator composed of ultralight silicon. Kang et. al. (2019) proposed a lightweight hand-assistive device made of only polymer materials to assist the tendon drive. These technologies have been developed for people with disabilities who can grasp objects with their hands without completely bending their fingers. Moreover, Wang et. al. (2017) designed a rehabilitation exercise glove that repeats finger flexion and extension motions using a lightweight bidirectional pneumatic actuator. Cappello et. al. (2018) developed a fabric-based finger-muscle strength-assisting glove that is inserted using a pneumatic method, and they succeeded in manipulating light objects with a force that was approximately 30% of the maximum gripping power of adults. Ge et. al. (2020) developed rehabilitation-exercise gloves by designing a pneumatic actuator based on knitted textiles and fabrics. Thus, rehabilitation-assistive technology increases the usability of assistive devices by minimizing the burden on the human body through textile-based designs.
Shape-memory alloy and its applications
Among SMAs, Nitinol (Ni–Ti), an alloy of nickel and titanium, is the most used alloy that exhibits a shape-memory effect, and it can exhibit a contraction of up to 8% under thermal stimulation (Eschen et al., 2020). SMAs with these shape-transforming properties have been used in various fields such as mechanical engineering, aviation, and protective clothing. Holschuh and Newman (2014) designed a spiral SMA wire in tiny microunits for potential applicability in the aerospace field. Wang et. al. (2020) demonstrated that the air gap between the fabric layers created by an SMA spring can be applied to temperature-sensitive protective clothing. Ozbulut et. al. (2016) designed an SMA cable composed of multiple layers of wire and demonstrated that a high driving speed can be achieved with less energy. Furthermore, some attempts have been made to maximize the transformation of SMA materials by designing new structures, such as cables or spiral shapes. Nonetheless, such efforts have a limited impact on wearable devices owing to the reduced flexibility resulting from the structures added to the actuator.
As an alternative solution to the soft wearable researcher’s dilemma between maximizing the SMA actuation performance and improving wearability, some researchers suggested SMA wires in a knit structure as it provided flexibility as well as a greater actuation force relative to the amount of material used. For example, Han and Ahn (2017) presented various 4D deformations through the knitted SMA, and compared the loads according to each pattern. Kim and Kim (2022) compared and analyzed the actuating force of a manufactured knitted SMA module through simulation using finite element analysis. Furthermore, knitted SMA can potentially be applied to soft wearable products because it exhibits more than 40% improvement in contractile force, higher energy density, and improved mechanical performance compared to those of SMA wires, which present the unique flexible characteristics of knitted textiles (e.g., Eschen et al., 2020). Granberry et. al. (2019) developed a knitted SMA actuator for a calf-swelling preventive compression product that fits along the curve of the human body. Lee et. al. (2022a) suggested the possibility of an exoglove application by comparing and analyzing the joint finger ROMs of older adults and applying the bending motion of the knitted SMA.
Methods
Participants
Adopting the multiple-case study approach, we recruited the three hemiplegic patients from a local community rehabilitation center, located in a metropolitan area, as the three individual use cases. According to Yin (1981, 2002), the multiple-case study approach, which is suitable to apply for hard-to-reach participants (e.g., hemiplegic patients), offers an effective venue to collect sufficient information from several different sources (or individuals), enabling to reach a sound conclusion. As summarized in Table 2, the three hemiplegic patients recruited for this study showed different demographic, anthropometric and clinical profiles (see Table 1), yet they all satisfied the following criteria: (a) they were aged 60 years or older; (b) had a disability grade of 1+ or lower on the Modified Ashworth Scale (MAS); (c) could bend their finger within less than half of their finger joint ROM without resistance; and (d) had experienced limited physical abilities in daily activities. Patients diagnosed with language, intellectual, or mental disorders were excluded from this study. We also recruited a physical therapist with 15 years of experience in rehabilitation therapy who served as a senior physical therapist at the center. The physical therapist provided expert opinions on the rehabilitation needs of hemiplegic patients. All participants signed the written informed consent before participating in the study. This study was approved by the Institutional Review Board Committee of the participating university (IRB No. 2112/002-019).
Procedures
User needs identification
We conducted semi-structured interviews with participants to identify specific user needs. The interview questions for the hemiplegic patients consisted of 18 items in total (Table 2). The questions inquired about the patients’ general behaviors for rehabilitation (nine items), previous use experience with wearable assistive devices (four items), and personal preferences for the devices (five items). The interview questions for the physical therapist included nine questions concerning medical symptoms and rehabilitation methods for hemiplegic patients (four items) and functional requirements for new wearable assistive devices (five items). Following Lamb and Kallal’s (1992) FEA consumer needs model, interview data were analyzed for functional, expressive, and aesthetic needs (refer to Table 4 for results).
Fabrication of a knitted SMA actuator
By adopting Han and Ahn (2017)’s approach, we fabricated an actuator using an SMA wire wrapped in a thread (100% polyester), which was originally coined to prevent short-circuit leakage between loops and increase the friction coefficient during the deformation phase of the SMA wire (Table 3). We used a plain knit having unique bending characteristics in that the needle loops in each loop are bent toward the leg loop (Lee et al., 2022a). We also used double modules in which a plain-knit and rib-knit overlapped because they contract in the course direction during actuation (Jung et al., 2022). Additionally, we measured the forces and bending angles of the plain-knit and double-knit modules to analyze the performance of the knitted SMA (Fig. 1). The actuation force was measured using a multiaxis force sensor (F/T Sensor-Nano 17, ATI Industrial Automation, USA), and the bending angle was measured over time using an online protractor on the knit-module side (Iconico Inc., USA) from the start point to the end point of the deformation. The knit module was divided into three parts—top, middle, and bottom, and the angle at each of these three points was defined as the bending angle.
Exoglove design and prototyping
We referred to the specific user needs identified in the previous stage, and decided on the overall structure and design details to meet the users’ functional, expressive and aesthetic needs. Additionally, we determine a specific body part to attach an SMA actuator on the exoglove based on the user’s hand dimensions and joint movement. We obtained the 3D shapes of the hemiplegic patients’ hands in a relaxed posture using a 3D body scanner (Artec Eva, Artec 3D, Luxembourg) and used the 3D data to draft the glove patterns (Lee et al., 2022b). We first 3D-scanned the hemiplegic patients’ hands, filled the missing holes on the 3D hand scan, flattened the surface of the 3D hand scan into 2D patterns in reverse engineering software (Geomagic Design X, 3D Systems, USA) (refer to Table 1), and finally customized the glove patterns to fit the hand of the individual patient. We also optimized the driving force of the knitted SMA actuator in the joint areas through an iterative prototyping process. The knitted SMA modules were then attached to the developed glove and positioned at each finger joint to assist with force when performing the hand extension exercise.
Evaluation
For evaluation of the prototyped exoglove, we carried out tests to gauge the functional performance of the knitted SMA and wearability of the exoglove. For testing of functional performance, we measured the joint ROM changes before and after glove actuation using a digital goniometer (EasyAngle, Meloq, Sweden). Subsequently, we analyzed the differences in the joint ROM and grip force before and after the four stimuli using a hand dynamometer (FBA_EH101, Camry, USA) to confirm the multiple effects of thermal stimulation and exercise assistance using the developed glove. The four stimuli were as follows: (a) finger extension exercise for 20 times (S1); (b) thermal stimulation using SMA actuation for 60 s (S2); (c) S1 and S2 simultaneously (S3); and (d) S3 and prototyped glove actuation simultaneously (S4). A baseline with no stimulus was denoted as S0. All evaluations were performed in triplicate.
To evaluate the wearability of the exoglove, we asked the hemiplegic patients to perform donning and doffing of the prototyped glove three times each and complete the Quebec user evaluation of satisfaction with assistive technology (QUEST 2.0) on a five-point Likert scale (Demers et al., 2002). Of the 12 items in QUEST 2.0, we retrieved the eight items concerning assistive devices and used them for subjective evaluation. Furthermore, to ensure user safety from the Joule heat from the SMA actuation, we analyzed changes in the skin temperature of the patients’ hands, using the data measured for 1 min using the iButton (DS1923, Dallas Instruments, USA), and assessed the patients’ subjective thermal perception and comfort following the ISO standard (ISO 10551, 2019). Thermal perception, which is perceived thermal sensation, was measured on a 9-point Likert scale (− 4 = very cold; 0 = neutral; 4 = very hot), and thermal comfort, the pleasantness of thermal sensation, was assessed using a 7-point Likert scale (− 3 = very unpleasant; 0 = neutral; 3 = very comfortable).
Data analysis
We used SPSS ver. 26.0 (IBM SPSS Statistics 26, USA) to analyze the joint ROM, grip force, and evaluation data. Because of the small sample size, the normality of the data was not obtained based on the Shapiro–Wilk test results (p > 0.050). Therefore, we performed nonparametric statistical analyses, i.e., the Kruskal–Wallis H tests, for the mean comparison in the joint ROMs; and Mann–Whitney U tests for the differences between knitted SMA modules. Additionally, we conducted one-way ANOVA and post-hoc tests to determine the mean differences in joint ROMs and grip forces across the four stimuli conditions (S0–S4). Statistical significance was set at a 95% confidence level for all analyses.
Results and Discussion
Identified user needs
Table 4 summarizes the user needs identified from the interviews with the participating hemiplegic patients and a physical therapist. As for the functional needs, the patients indicated difficulties in wearing gloves due to stiffness, understanding the use of new devices, and problems with low skin temperature on the hemiplegic side. There were needs for ease of donning/doffing, comfort, ease of movement during wearing gloves, and a heating function for the hemiplegic side and home care. A physical therapist pointed out muscle contraction, nerve compression, and wrist drop of the patients’ hands as the functional considerations. The patients' need for safety while using such wearable devices was also noted. Regarding the expressive needs, the patients had been reluctant to be seen as disabled or patients by wearing assistive devices, and there was a need to be seen as an active and confident person. Similarly, the physical therapist indicated a need for familiarity as patients had problems with reluctance to use new rehabilitation equipment. As for the aesthetic needs, the patients were hesitant to use inconvenient and standing out designs. Thus, details should be minimized, and an inconspicuous color needs to be used. A physical therapist raised a problem with the machine-like design of existing assistive devices; therefore, there was a need to minimize the exposure of driving parts. Although alleviating the symptoms of hemiplegic patients was an essential factor, other unsatisfactory factors could cause low usability (Bettoni et al., 2016). In other words, from the FEA analysis, we noted key user needs for aesthetic and expressive design as well as functionalelements in developing exogloves. This supports the results of previous studies (Nasir & Troynikov, 2019).
Knitted SMA actuator
To select the knitted SMA modules suitable for each finger joint and apply them as an actuator, the characteristics of SMA plain- and double-knit modules were identified. In the plain- and double-knit modules, the actuation force increased until approximately 10 s after the start of actuation, and it was then maintained at a constant (Fig. 2a). The phase changes of the plain-knit and double-knit modules started after 10 s and before 10 s and ended after 20 s and 25 s, respectively. The double-knit module had a longer phase-change time because the two modules were driven. In addition, the bending angle of the plain-knit module increased until 10 s, but became constant after 14 s, and the bending transformation no longer occurred (Fig. 2b). The angle of the double-knit module exhibited a rapidly increasing trend until approximately 7 s and then became constant after 16 s.
According to the Mann–Whitney U test, the actuation force and bending angle tended to be higher in the plain-knit than in the double knit; however, the difference was not statistically significant (all p = 0.050) (Table 5). The electrical power values applied during actuation were 0.18 (0.01) and 0.29 (0.02) Wh for the plain- and double-knit modules, respectively. Although the difference in electrical power between the two modules was not significant (p = 0.513), less power tended to be consumed during the plain-knit operation. Based on these results, plain-knit modules with higher bending angles and actuation forces were applied to the distal interphalangeal (DIP) and proximal interphalangeal (PIP) joints, where the joint ROM of the patients tended to be smaller than those of other joints. In contrast, double-knit modules with a relatively small actuation force and bending angle were applied to the metacarpal phalangeal (MCP) joint, which showed a greater joint ROM than those of the other joints (Fig. 3). These findings corresponded to the results of a previous study (Lee et al., 2023) that proposed design guidelines for the knitted SMA for assistive wearable applications based on human data from hemiplegic patients.
Exoglove design and prototyping
As specific design solutions to satisfy the functional needs, we included the open-type enclosure system in the exoglove for ease donning and doffing. For the comfort and ease of movement, elastic materials were used for the joints and between the fingers with a large ROM. We attempted to provide thermal stimulation using the Joule heat generated when operating the SMA to relax stiff muscles. For ease of operation, the power controller was designed using an on/off actuating method that operated with a single button. A 2.50-mm-thick double-sided neoprene material was used in the wrist area to minimize the wrist-drop phenomenon. To respond to the expressive needs, we tried to show the wearers as confident and active human-beings and did not emphasize their disabilities. To minimize the fear of using new devices and familiarizing the users, the device was designed in the same shape as regular thermal gloves. This design minimizes the gaze of others. In addition, the operating part of the glove was fabricated using a knitted SMA and covered with fabric to make it discrete. For the aesthetic needs, light, soft-touch, and stretchable materials mixed with polyurethane fibers were used, except at the wrist area. Additionally, lightweight and flexible nylon materials were primarily used for the joints. For an inconspicuous design, achromatic colors such as gray and black were used in keeping with patients’ comments.
We developed a soft wearable exoglove through the iterative prototyping process. To draft the ergonomic patterns of the designed exoglove, the glove pattern of each patient was designed by referring to the glove-pattern design method (Lee et al., 2022b). The developed exoglove comprised the four layers, as seen in Fig. 4. The innermost part that encountered the skin was designed with an insulation fabric for safety against the Joule heat generated during the actuation of the SMA modules. In the second layer, an elastic material was used for the joint area as the base fabric to which the actuator was attached to provide comfort during hand movement. For the wrist area, double neoprene material was used to minimize wrist drops on the hemiplegic side. Fourteen knitted SMA modules were attached to the base fabric. A porous material was used on the outer part of the glove to cover the SMA modules so that they were not visible from the outside and to dissipate heat. In addition, an Arduino AT mega 2560 R3 microcontroller-compatible board and a power supply were used to operate the SMA module.
Evaluation on functional performance
Changes in joint ROMs
The results of Kruskal–Wallis H tests for differences before and after wearing exoglove and after actuating the SMA were as follows. After actuating the developed exoglove, the joint ROM of the patients’ fingers increased significantly to 1.72° (13.71%) (p = 0.006). The joint ROM exhibited an increasing trend at each joint, and the increase rate was the largest (18.95%) at the PIP joint (p = 0.008). Although the joint ROM of the DIP and MCP joints increased by 10.56% and 11.66%, respectively, these differences were not statistically significant (p = 0.170 and 0.118, respectively). Furthermore, the joint ROMs were significantly different across the stimuli at the 95% confidence level (Table 6), which were in the following order from the highest to the lowest: S4 (35.61 ± 15.88°) > S3 (30.35 ± 12.17°) > S2 (23.88 ± 11.68°) > S1 (14.78 ± 10.90°). The post hoc test result (Table 7) also showed mean comparisons between the stimuli. Specifically, the joint ROMs increased after each stimulation of S2–S4 (all p = 0.000) from S0; however, they were not significant between S0 and S1 (p = 0.943).
Changes in grip strength
As seen in Table 6, the result of the ANOVA test determined that the grip forces were significantly different across the stimuli (p = 0.000). In particular, the grip forces were in the following order from the highest to the lowest: S4 (6.96 ± 4.20 kg) > S3 (6.32 ± 3.98 kg) > S2 (5.78 ± 3.73 kg) > S1 (5.25 ± 3.35 kg). Further, as seen in Table 7, the results of the post hoc test revealed that the grip forces noticeably increased after each stimulation of S2–S4 at the 95% confidence level (p = 0.024, 0.011, and 0.003, respectively), as compared to S0, but not S1 (p = 0.109), which signified that all other stimuli conditions, except S1, were effective in improving the participants’ grip forces.
Wearability assessment
Donning and doffing times
The average donning and doffing times of the hemiplegic patients were 58.81 s (SD 14.70) and 8.67 s (SD 3.57), respectively; thus, more time was required for donning than for doffing (p = 0.008). The main donning and doffing processes could be divided into four tasks (Table 8). Specifically, the donning task consisted of, in order, inserting the four fingers into the glove, inserting the thumb, holding a wrist Velcro with the opposite hand on the hemiplegic side, and closing the Velcro. The most time was required to insert the thumb and four fingers into the glove (p = 0.000). The doffing task was performed as follows: gripping and removing the wrist Velcro with the opposite hand on the hemiplegic side, holding the tips of the four fingers, pulling the glove, and doffing the fingers from the gloves. A considerable amount of time was required to grip the Velcro, hold the tip of the glove, pull the fingertips, and doff the glove.
Subjective satisfaction
All eight items scored higher than average. In six of the eight items, more than 4.00 points (“satisfaction”) were scored on the 5-point Likert scale, and safety and durability items scored 3.67 points (between “normal” and “satisfaction”). The highest score was 4.33 for size and adjustments, and the lowest score was 3.67 for safety and durability. The weight, ease of use, comfort, and effectiveness items earned a “satisfaction” score of 4.00 points. In addition, the two items that patients considered most important were size and comfort. This is because hemiplegic patients were not typically able to use gloves owing to their rigid and bent hands as well as the related discomfort.
Ensured thermal safety
The average skin temperature of the patients’ hands when wearing the exoglove was 30.35 °C (SD 0.57). After actuating the glove, the skin temperature increased from 30.32 °C (SD 0.30) to 35.20 °C (SD 0.72) in 1 min. The average difference in skin temperature between that while wearing the glove and that after glove actuation was statistically significant (p = 0.000). Moreover, the average temperature increased by 2.21 °C for 60 s after actuation compared to that before. For the subjective thermal evaluation, the average score for thermal perception was 1.00 (SD 0.82), which was evaluated as “slightly warm.” These thermal perceptions are an evaluation of the inside of the glove, that is, the part of the hand on which the glove was worn during actuation. All patients positively evaluated the increase in skin temperature on the hemiplegic side because they tended to have lower skin temperatures on the hemiplegic side than those on the opposite side (Matsumoto et al., 2010). Patients felt that the hand on the hemiplegic side warmed while performing the extension exercise, as the temperature gradually increased from the start of glove actuation. The subjective thermal perception was evaluated as “slightly warm,” with a score of 1.00 (SD 0.82). This result showed that the temperature that gradually increased from the start of the glove drive warmed the hand on the hemiplegic side while performing the extension movement. Thermal comfort was positively evaluated between “comfortable” and “slightly comfortable,” with a score of 0.67 (SD 0.47). The participants gave positive evaluations beyond “comfort” regarding the provision of thermal stimulation while using the developed gloves. These findings, as a psychological indicator of heat, were able to support the positive influence of the thermal stimulation of the developed gloves through the actually measured quantitative skin temperature change. Thus, these results support those of previous research that have suggested positive effects, such as increased blood circulation through local thermal stimulation (Matsumoto et al., 2010). Furthermore, these results showed that the immediate effect on joint ROMs and gripping force increase for a combined effect was greater than when only the extension exercise or thermal stimulation was applied (Wang et al., 2017; Yoon et al., 2018).
Conclusions
This study developed hand rehabilitation assistive gloves for hemiplegic patients, using the knitted SMA as a flexible actuator. To outline the outcomes of the study, we identified the patients’ key problems through the systematic FEA needs analysis, which were then applied in designing a customizable glove for an individual patient, along with the hand anthropometric information. Following, we determined the positive effects of the developed gloves on the hand ROM, bending performance, and gripping force of the patients, as well as wearability. Unlike the previous studies on the SMA that mainly focused on characterizing the actuating performance (e.g., Eschen et al., 2020; Han & Ahn, 2017), this study validated the potential of the smart material’s suitability for wearable robotics that are intended to be worn on the human body, especially for rehabilitation, which offers a significance of the present study.
Nonetheless, this study had some limitations. While the multiple case study approach offers a strong advantage for an in-depth investigation (Schoch, 2020), additional empirical evidence is desirable to be generalizable in other contexts (Baxter & Jack, 2008). Although three hemiplegic patients with MAS grade 1 who exhibited different use cases showed positive results, the effectiveness of the gloves may vary when applied to those with different clinical backgrounds. In addition, because a geographical location for patient recruitment was limited to one region and country, further research is appropriate with patients with diverse geographic and cultural profiles. Moreover, the actuating force of a knitted SMA varies depending on the accompanying components used in soft wearable robotics. Thus, additional research is desirable to minimize the size and weight of the actuator to make it more compatible with soft wearable robots. Future research could be focused on long-term actuation repeatability for SMA actuators and the material durability of rehabilitation gloves in terms of usability. Besides, we used nonparametric methods for data comparison to minimize the probability of Type II errors when the data is not normalized (Qualls et al., 2010). Despite the advantage of nonparametric methods, the statistical results should be carefully interpreted, as they tend to be less sensitive to detect significant differences, as compared to parametric tests (Campbell, 2021). However, given the difficulty of obtaining immediately normal data in experimental research, we propose to accumulate and validate a larger sample set with statistical confidence. All in all, we conclude that a knitted SMA is a smart fabric-structured material promising for use as an effective actuator with shape transformation and flexibility, and our soft wearable exoglove integrated with the knitted SMA-based actuator presented positive benefits for rehabilitation assistance with good wearability.
References
Baxter, P., & Jack, S. (2008). Qualitative case study methodology: Study design and implementation for novice researchers. The Qualitative Report, 13(4), 544–559.
Bettoni, E., Ferriero, G., Bakhsh, H., Bravini, E., Massazza, G., & Franchignoni, F. (2016). A systematic review of questionnaires to assess patient satisfaction with limb orthoses. Prosthetics and Orthotics International, 40(2), 158–169. https://doi.org/10.1177/0309364614556836
Campbell, M. J. (2021). Statistics at square one (12th ed.). Wiley.
Cappello, L., Meyer, J. T., Galloway, K. C., Peisner, J. D., Granberry, R., Wagner, D. A., Engelhardt, S., Paganoni, S., & Walsh, C. J. (2018). Assisting hand function after spinal cord injury with a fabric-based soft robotic glove. Journal of Neuroengineering and Rehabilitation, 15(1), 1–10. https://doi.org/10.1186/s12984-018-0391-x
Cauraugh, J., Light, K., Kim, S., Thigpen, M., & Behrman, A. (2000). Chronic motor dysfunction after stroke: Recovering wrist and finger extension by electromyography-triggered neuromuscular stimulation. Stroke, 31(6), 1360–1364. https://doi.org/10.1161/01.STR.31.6.1360
De Witte, L., Steel, E., Gupta, S., Ramos, V. D., & Roentgen, U. (2018). Assistive technology provision: Towards an international framework for assuring availability and accessibility of affordable high-quality assistive technology. Disability and Rehabilitation: Assistive Technology, 13(5), 467–472. https://doi.org/10.1080/17483107.2018.1470264
Demers, L., Weiss-Lambrou, R., & Ska, B. (2002). The Quebec user evaluation of satisfaction with assistive technology (QUEST 2.0): An overview and recent progress. Technology and Disability, 14(3), 101–105. https://doi.org/10.13072/midss.298
Eschen, K., Granberry, R., & Abel, J. (2020). Guidelines on the design, characterization, and operation of shape memory alloy knitted actuators. Smart Materials and Structures, 29(3), 035036. https://doi.org/10.1088/1361-665X/ab6ba7
Ge, L., Chen, F., Wang, D., Zhang, Y., Han, D., Wang, T., & Gu, G. (2020). Design, modeling, and evaluation of fabric-based pneumatic actuators for soft wearable assistive gloves. Soft Robotics, 7(5), 583–596. https://doi.org/10.1089/soro.2019.0105
Granberry, R., Eschen, K., Holschuh, B., & Abel, J. (2019). Functionally graded knitted actuators with NiTi-based shape memory alloys for topographically self-fitting wearables. Advanced Materials Technologies, 4(11), Article 1900548. https://doi.org/10.1002/admt.201900548
Hadi, A., Alipour, K., Kazeminasab, S., & Elahinia, M. (2018). ASR glove: A wearable glove for hand assistance and rehabilitation using shape memory alloys. Journal of Intelligent Material Systems and Structures, 29(8), 1575–1585. https://doi.org/10.1177/1045389X17742729
Han, M. W., & Ahn, S. H. (2017). Blooming knit flowers: Loop-linked soft morphing structures for soft robotics. Advanced Materials, 29(13), Article 1606580. https://doi.org/10.1002/adma.201606580
Holschuh, B., & Newman, D. (2014). Low spring index, large displacement shape memory alloy (SMA) coil actuators for use in macro-and micro-systems. Proceedings of Reliability, Packaging, Testing, and Characterization of MOEMS/MEMS, Nanodevices, and Nanomaterials XIII, 8975, 23–33. https://doi.org/10.1117/12.2044406
In, H., Kang, B. B., Sin, M., & Cho, K. J. (2015). Exo-glove: A wearable robot for the hand with a soft tendon routing system. IEEE Robotics & Automation Magazine, 22(1), 97–105. https://doi.org/10.1109/MRA.2014.2362863
International Organization for Standardization. (2019). Ergonomics of the physical environment—Subjective judgement scales for assessing physical environments. (ISO Standard No. 10551:2019).
Ji, J. G., Park, C. H., & Kauk, Y. S. (2013). The structural relationships among autonomy support, psychological needs, and movement behavior of high-school team sports players. Journal of Coaching Development, 15(4), 173–181.
Jung, W. K., Lee, S. M., Ahn, S. H., & Park, J. (2022). Development and assessment of a knitted shape memory alloy-based multifunctional elbow brace. Journal of Industrial Textiles, 51(2S), 1989S-2009S. https://doi.org/10.1177/15280837211056983
Kang, B. B., Choi, H., Lee, H., & Cho, K. J. (2019). Exo-glove poly II: A polymer-based soft wearable robot for the hand with a tendon-driven actuation system. Soft Robotics, 6(2), 214–227. https://doi.org/10.1089/soro.2018.0006
Kawahira, K., Shimodozono, M., Etoh, S., Kamada, K., Noma, T., & Tanaka, N. (2010). Effects of intensive repetition of a new facilitation technique on motor functional recovery of the hemiplegic upper limb and hand. Brain Injury, 24(10), 1202–1213. https://doi.org/10.3109/02699052.2010.506855
Kim, J., Quinlivan, B. T., Deprey, L. A., Arumukhom Revi, D., Eckert-Erdheim, A., Murphy, P., Orzel, D., & Walsh, C. J. (2022). Reducing the energy cost of walking with low assistance levels through optimized hip flexion assistance from a soft exosuit. Scientific Reports, 12(1), Article 11004. https://doi.org/10.1038/s41598-022-14784-9
Kim, S., & Kim, J. (2022). Analysis of driving forces of 3D knitted shape memory textile actuators using scale-up finite element method. Fashion and Textiles, 9(1), Article 38. https://doi.org/10.1186/s40691-022-00307-7
Kim, S. B., Kim, Y. J., Kim, C., & Park, M. S. (2005). The effect of heat therapy on cutaneous blood flow and skin temperature at pre-auricular region. Journal of Oral Medicine and Pain, 30(4), 401–410.
Lamb, J. M., & Kallal, M. J. (1992). A conceptual framework for apparel design. Clothing and Textiles Research Journal, 10(2), 42–47. https://doi.org/10.1177/0887302X9201000207
Lee, S. M., Jung, W. K., & Park, J. (2023). Mapping actuating performance of knitted shape memory alloys and proposing design guidelines for assistive wearable applications. Journal of Industrial Textiles, 53, 1–25. https://doi.org/10.1177/15280837231211001
Lee, S. M., Jung, W. K., Park, J., & Ahn, S. H. (2022a). Development of a 4D hand gripping aid using a knitted shape memory alloy and evaluation of finger-bending angles in elderly women. Fashion and Textiles, 9(1), Article 11. https://doi.org/10.1186/s40691-021-00280-7
Lee, S. M., Kim, M. K., & Park, J. (2022b). An ergonomic glove pattern drafting method for hand assistive devices. In Proceedings of the international textile and apparel association annual conference, 77(1). Iowa State University Digital Press.
Matsumoto, S., Shimodozono, M., Etoh, S., Shimozono, Y., Tanaka, N., & Kawahira, K. (2010). Beneficial effects of footbaths in controlling spasticity after stroke. International Journal of Biometeorology, 54(4), 465–473. https://doi.org/10.1007/s00484-009-0300-x
Nasir, S. H., & Troynikov, O. (2019). Therapeutic gloves for arthritis: Development of a design framework. International Journal of Fashion Design, Technology and Education, 12(3), 346–355. https://doi.org/10.1080/17543266.2019.1641845
Ozbulut, O. E., Daghash, S., & Sherif, M. M. (2016). Shape memory alloy cables for structural applications. Journal of Materials in Civil Engineering, 28(4), Article 04015176. https://doi.org/10.1061/(ASCE)MT.1943-5533.0001457
Qualls, M., Pallin, D. J., & Schuur, J. D. (2010). Parametric versus nonparametric statistical tests: The length of stay example. Academic Emergency Medicine, 17(10), 1113–1121.
Quinlivan, B. T., Lee, S., Malcolm, P., Rossi, D. M., Grimmer, M., Siviy, C., Galiana, I., & Walsh, C. J. (2017). Assistance magnitude versus metabolic cost reductions for a tethered multiarticular soft exosuit. Science Robotics, 2(2), Article eaah4416. https://doi.org/10.1126/scirobotics.aah4416
Schoch, K., et al. (2020). Case study research. In Gary (Ed.), Research design and methods: An applied guide for the scholar-practitioner (pp. 245–258). SAGE Publications.
United Nations. (2022). Ageing and disability. United Nations Department of Economic and Social Affairs. Retrieved February 19, 2024, from https://www.un.org/development/desa/disabilities/disability-and-ageing.html
Walsh, C. (2018). Human-in-the-loop development of soft wearable robots. Nature Reviews Materials, 3(6), 78–80. https://doi.org/10.1038/s41578-018-0011-1
Wang, B., McDaid, A., Aw, K. C., & Biglari-Abhari, M. (2017). Design and development of a skinny bidirectional soft glove for post-stroke hand rehabilitation. In Proceedings of the 2017 intelligent systems conference (IntelliSys), pp 979–987. https://doi.org/10.1109/IntelliSys.2017.8324248.
Wang, L., Lu, Y., & He, J. (2020). On the effectiveness of temperature-responsive protective fabric incorporated with shape memory alloy (SMA) under radiant heat exposure. Clothing and Textiles Research Journal, 38(3), 212–224. https://doi.org/10.1177/0887302X19892095
Wang, Y., Zheng, S., Pang, J., Li, S., & Li, J. (2021). Design and experiment of a hand movement device driven by shape memory alloy wires. Journal of Robotics, 2021, 1–13. https://doi.org/10.1155/2021/6611581
World Health Organization. (2023, March 7). Disability. https://www.who.int/news-room/fact-sheets/detail/disability-and-health.
Yeung, K. T., Lin, C. H., Teng, Y. L., Chen, F. F., Lou, S. Z., & Chen, C. L. (2016). Use of and self-perceived need for assistive devices in individuals with disabilities in Taiwan. PLoS ONE, 11(3), Article e0152707. https://doi.org/10.1371/journal.pone.0152707
Yin, R. K. (1981). The case study as a serious research strategy. Knowledge, 3(1), 97–114.
Yin, R. K. (2002). Case study research: Design and methods. SAGE Publications.
Yoon, J. S., Kim, Y. C., Kim, M. J., & Kim, J. H. (2018). An air pocket glove for finger rehabilitation and quantitative assessment of hemiplegic patients. Journal of the Korean Society for Precision Engineering, 35(8), 817–823. https://doi.org/10.7736/KSPE.2018.35.8.817
Zhang, Q., Wommer, J., O’Rourke, C., Teitelman, J., Tang, Y., Robison, J., Lin, G., & Yin, J. (2017). Origami and kirigami inspired self-folding for programming three-dimensional shape shifting of polymer sheets with light. Extreme Mechanics Letters, 11, 111–120. https://doi.org/10.1016/j.eml.2016.08.004
Acknowledgements
Not applicable.
Funding
This study was supported by the National Research Foundation of Korea (NRF) grant funded by the Korean Government (MSIT) [RS-2023-00208052]; the Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education [2020R1A6A3A01099046], and Alchemist Project funded by the Korean Ministry of Trade, Industry and Energy [20007058-K_G012000705804-10054408].
Author information
Authors and Affiliations
Contributions
All authors collaboratively contributed to the design, planning, and execution of the research project. SL as the first authors were responsible for collecting and analysing the data and writing the first draft of the manuscript. JP supervised the overall research process and reviewed and revised the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare that there is no conflict of interest.
Ethical approval and informed consent
This research was conducted under the approval and supervision of Seoul National University Institutional Review Board (IRB Approval No: 2112/002-019) regarding ethical issues including consent to participate.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Lee, SM., Park, J. A soft wearable exoglove for rehabilitation assistance: a novel application of knitted shape-memory alloy as a flexible actuator. Fash Text 11, 14 (2024). https://doi.org/10.1186/s40691-024-00377-9
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s40691-024-00377-9